B6.3 - Field Guide to Nonlinear Calibration of Fiber Bragg Gratings for Temperature Sensors with a Wide Temperature Range

Event
AMA Conferences 2017
2017-05-30 - 2017-06-01
Nürnberg, Germany
Band
Proceedings Sensor 2017
Chapter
B6 - Distributed and Fiber Bragg Grating Sending I
Author(s)
T. Ringel, T. Bosselmann - Siemens AG, Erlangen (Germany)
Pages
267 - 270
DOI
10.5162/sensor2017/B6.3
ISBN
978-3-9816876-4-4
Price
free

Abstract

This paper presents a field guide to nonlinear calibration of Fiber Bragg Gratings (FBG1,2) for temperature sensors with a wide temperature range. An FBG which is used for a wide temperature range needs an expensive calibration curve measured for this particular FBG to enable the calculation of the temperature from the measured wavelengths. We show here that it is possible to get the calibration curve for any FBG out of the measurement of a few FBGs and an inexpensive measurement of the wavelength of this FBG e.g. at 0°C. Zero degree is preferred because that can be produced by ice water, for other temperatures a calibration oven is needed.

Download