Banner

T5.2.3 - Signal Averaging with Heartbeat Envelope for Non-Contact Heart Sound Measurement

Event
EUROSENSORS XXXVII
2025-09-07 - 2025-09-10
Wroclaw
Band
Lectures
Chapter
T5.2 - Sensors for Medicine and Life-Sciences 1
Author(s)
S. Muramatsu, J. W. Fastier-Wooller, M. Yamamoto, T. Itoh - The University of Tokyo, Tokyo (Japan)
Pages
134 - 135
DOI
10.5162/EUROSENSORS2025/T5.2.3
ISBN
978-3-910600-07-2
Price
free

Abstract

We propose a non-contact heart sound measurement method using signal averaging in conjunction withthe heartbeat envelope as a trigger. The performance of the heartbeat envelope was 0.01 s as thestandard deviation of the R-R interval estimation error, meeting the trigger requirement for our signal averaging method. The main component of the heart sound was extracted from a sound signal meas-ured 40 mm away from the chest wall.