Banner

B7-b3 - A 128-channel Ultrasound Imaging System for High Frame Rate Imaging

Event
2025 ICU PADERBORN - 9th International Congress on Ultrasonics
2025-09-21 - 2025-09-25
Paderborn
Band
Lectures
Chapter
B7-a - High-Frequency Ultrasound Imaging
Author(s)
X. Zhang, J. Ma - Beihang University, Beijing (China)
Pages
157 - 160
DOI
10.5162/Ultrasonic2025/B7-b3
ISBN
978-3-910600-08-9
Price
free

Abstract

Ultrafast ultrasound imaging has opened up new avenues for applications in clinical diagnostics and biomedical research. We developed an open ultrasound platform based on a single FPGA to enable high-frame-rate beamforming. The system comprises a commercial PC for user interaction and an ultrasound acquisition card connected via PCIe 3.0, supporting real-time beamforming and programmable transmission/reception. The parallel delay-and-sum beamformer is implemented using data register shifting technology and can output beamformed images of 128×2048 at a frame rate of 10,557 fps.