D2-3.2 - Identification of Bacteria by Pyrolysis-Gas Chromatography-Ion Mobility Spectrometry

Event
23. ITG/GMA-Fachtagung Sensoren und Messsysteme 2026
2026-06-09 - 2026-06-10
Nürnberg
Band
Vorträge
Chapter
Biosensoren
Author(s)
D. Röckrath, M. Hitzemann, A. Höhne, P. Timmermann, S. Zimmermann - Leibniz University Hannover, Hannover, A. Vargas-Valderrama - Royal Military Academy, Brussels (Belgium)
Pages
224 - 225
DOI
10.5162/sensoren2026/D2-3.2
ISBN
978-3-910600-11-9
Price
free

Abstract

This work presents a novel approach for rapid and reliable analysis of biological samples, crucial in medical, safety, and security sectors. The self-built device combines a pyrolyzer with a gas chromatograph and an ultra-fast ion mobility spectrometer. This portable system allows sensitive analysis of liquid and solid samples on-site, including viruses and bacteria. The sample undergoes pyrolysis under an inert atmosphere, breaking down larger non-volatile compounds into their volatile constituents. The gas chromatograph separates the constituents, while the ion mobility spectrometer provides an orthogonal dimension of separation and detection at the highest sensitivity. The method effectively differentiates between bacteria such as Streptococcus Mitis and Streptococcus Suis.