1.3.2 - Continuous Respiratory Rate Monitoring Using a Flexible Thermoelectric Sensor

Event
EUROSENSORS 2026
2026-09-06 - 2026-09-09
Zurich
Band
Lectures
Chapter
Sensing for Medical Diagnostics
Author(s)
A. Johnson-Brett, A. Angelopoulos, R. Zafeirelis, H. Akerman, I. Zeimpekis, N. Harris, N. White, K. A. Morgan - University of Southampton,Southampton (United Kingdom)
Pages
58 - 59
DOI
10.5162/eurosensors2026/1.3.2
ISBN
978-3-910600-12-6
Price
free

Abstract

Respiratory rate is an early indication of patient decline. A key limitation of continuous respiratory moni-toring for hospital use is providing a non-invasive, low cost and easy to use system. Wearable sensors offer a continuous, lightweight and an unobtrusive solution. The proposed system combines a thin flexi-ble thermoelectric sensor into a standard hospital grade oxygen mask with continuous Bluetooth con-nectivity and a selected algorithm. The proposed system is compared to a medical standard with a bias of -0.04, limits of agreement of -0.34 and +0.26 breaths per minute and successful apnoeic detection.