1.3.3 - Comparative Assessment of Assimilating Probes and Molecular Beacons in Multiplex LAMP for Diagnostic

Event
EUROSENSORS 2026
2026-09-06 - 2026-09-09
Zurich
Band
Lectures
Chapter
Sensing for Medical Diagnostics
Author(s)
A. Mishra, T. L. Quyen, W. E. Svendsen, M. Dimaki - Technical University of Denmark,Kgs. Lyngby (Denmark)
Pages
60 - 61
DOI
10.5162/eurosensors2026/1.3.3
ISBN
978-3-910600-12-6
Price
free

Abstract

Despite widespread use of fluorescent probes in LAMP diagnostics, role of probe architecture in contributing to threshold time (Tt), independent of primer efficiency, remains unclear in multiplex assays. This gap limits rational probe selection for point-of-care applications. Here, we present a structure/function analysis framework comparing assimilating probes (APs) and molecular beacons (MBs) in multiplex LAMP targeting respiratory viruses. Briefly, APs out performed MBs in Tt and signal-to-background ratio for most of the targets in singleplex and multiplex reactions. These findings highlight the critical role of probe selection in developing multiplex LAMP assays for rapid diagnostics.