2.3.2 - Motion Artifact Removal for Moving FMCW Radar-Based Vital Sign Sensing Using IMU

Event
EUROSENSORS 2026
2026-09-06 - 2026-09-09
Zurich
Band
Lectures
Chapter
Sensing for Medical Diagnostics
Author(s)
N. Mascia, J. Fastier-Wooller, S. Muramatsu, M. Yamamoto, T. Itoh - The University of Tokyo,Tokyo (Japan)
Pages
94 - 95
DOI
10.5162/eurosensors2026/2.3.2
ISBN
978-3-910600-12-6
Price
free

Abstract

We propose an IMU-based motion artifact removal algorithm for non-contact vital sign monitoring, in which a frequency-modulated continuous-wave (FMCW) millimeter-wave (mm Wave) radar and an inertial measurement unit (IMU) are co-mounted on a moving platform. A least mean squares (LMS) adaptive filter exploits the correlation between IMU acceleration and radar phase disturbance to suppress motion-induced artifacts. Phantom experiments demonstrated a mean signal-to-noise ratio (SNR) improvement of 31.85 dB after calibration, while the cardiac signal at 1 Hz was preserved.