C1.2 A Radiation Thermometer based on an InGaAs-Photodiode at 1.6 ìm for Temperatures down to 80 °C

Event
SMSI 2020
-
(did not take place because of Covid-19 virus pandemic)
Band
SMSI 2020 - Sensors and Instrumentation
Chapter
C1 IRS² Satellite Conference: Infrared Detectors
Author(s)
J. Bories, I. Mueller, B. Gutschwager, J. Hollandt - Physikalisch-Technische Bundesanstalt, Berlin (Germany)
Pages
145 - 146
DOI
10.5162/SMSI2020/C1.2
ISBN
978-3-9819376-2-6
Price
free

Abstract

In the last decade the specific detectivity of InGaAs-photodiodes has improved significantly. Hence, the lower detection limit of InGaAs-photodiode based radiation thermometers at 1.6 μm can be extended. Here, an existing radiation thermometer was equipped with a state-of-the-art InGaAs-photodiode. The radiation thermometer was characterized and calibrated and a reference function has been compiled. The lower temperature limit could be extended from 150 °C down to 80 °C.