C3.1 Metrological Characterization and Calibration of Thermographic Cameras in the Temperature Range from 50 °C to 960 °C

Event
SMSI 2020
-
(did not take place because of Covid-19 virus pandemic)
Band
SMSI 2020 - Sensors and Instrumentation
Chapter
C3 IRS² Satellite Conference: Infrared Cameras and Thermal Imaging
Author(s)
S. König, B. Gutschwager, I. Müller, R. Taubert, J. Hollandt - Physikalisch-Technische Bundesanstalt, Berlin (Germany), F. Nagel - DIAS Infrared GmbH, Dresden (Germany)
Pages
161 - 162
DOI
10.5162/SMSI2020/C3.1
ISBN
978-3-9819376-2-6
Price
free

Abstract

We present the metrological characterization and calibration of four different types of thermographic cameras for quantitative temperature measurement traceable to the International Temperature Scale of 1990 (ITS-90). All measurements are performed according to the Technical Directives VDI/VDE 5585. Results achieved with different calibration procedures are compared for each type of camera and among the four cameras. An uncertainty budget for the calibration of each camera is set up ac-cording to the “Guide to the Expression of Uncertainty in Measurement” and VDI/VDE 5585 Part 2.