P2.7 Mode analysis for long, undamped cantilevers with added diamond tips of varying length for fast roughness measurements

Event
SMSI 2020
-
(did not take place because of Covid-19 virus pandemic)
Band
SMSI 2020 - Sensors and Instrumentation
Chapter
P2 Sensors and Measurement Systems for Physical Quantities
Author(s)
H. Behle, U. Brand - Physikalisch-Technische Bundesanstalt, Braunschweig (Germany)
Pages
238 - 239
DOI
10.5162/SMSI2020/P2.7
ISBN
978-3-9819376-2-6
Price
free

Abstract

In this article we present an FEM analysis of the change in resonance behaviour of 5 mm long piezoresistive cantilevers when they are modified with pencil shaped diamond tips of varying length. In a setup with glued cantilever base the length of the fixed bearing was examined as well. Simulations show that for a free cantilever base length of one fifth of the base no base oscillations occur. Base oscillations only occur for higher order modes of resonance with free base lengths greater one fifth.